The Elmetron CPC-511 is an easy-to-use laboratory meter for pH, mV (redox), conductivity up to 1000 mS/cm, salinity, TDS and temperature from -50.0 to 199.9 °C. It offers 1 to 3 point pH calibration with automatic buffer recognition and a large LCD display.
The Elmetron CPC-511 is an easy-to-use laboratory meter that measures pH, mV (redox potential), conductivity, salinity, TDS and temperature. Its large LCD shows the measured function and the temperature at the same time.
In the pH function the meter is calibrated in 1 to 3 points in buffer solutions, with automatic recognition of 4.00, 7.00 and 9.00 pH buffers (optionally 4.00, 7.00 and 10.00 pH), and it reports the condition of the pH electrode through the buffer and slope readouts. Depending on the electrode chosen, it measures in clean water, sewage, pastes and similar samples.
In the conductivity function it measures distilled water and other liquids up to 1000 mS/cm across 5 automatically switched sub-ranges, converts conductivity into salinity (NaCl up to 250 g/l or KCl up to 200 g/l) according to the real characteristic rather than a constant coefficient, and determines the estimated total dissolved solids (TDS). The temperature coefficient α is fixed at 2 %/°C; for precise temperature-compensated measurements Elmetron recommends its meters with a variable coefficient (CPC-502, CPC-505).
| Manufacturer | Elmetron |
|---|---|
| Model | CPC-511 |
| Measured parameters | pH, mV (redox potential), conductivity, salinity, TDS, temperature |
| pH range | -2.00 to 16.00 pH |
| pH accuracy | ±0.01 pH ±1 digit (meter only) |
| pH temperature compensation range | -5.0 to 110.0 °C |
| pH calibration | 1 to 3 points; automatic buffer detection 4.00, 7.00, 9.00 pH (optionally 4.00, 7.00, 10.00 pH) |
| mV range | ±1999 mV |
| mV accuracy | ±1 mV ±1 digit (meter only) |
| Input impedance (pH / mV) | 10¹² Ω |
| Conductivity range | 0 to 1000 mS/cm (autorange, 5 sub-ranges) |
| Conductivity accuracy | ±0.25 % ±1 digit (meter only) |
| Salinity range | up to 250 g/l NaCl; up to 200 g/l KCl |
| Salinity accuracy | 2 % ±1 digit (meter only) |
| Conductivity temperature compensation range | -5.0 to 70.0 °C |
| Temperature coefficient α | Constant: 2 %/°C |
| Cell constant K | 0.010 to 19.999 cm⁻¹ |
| Temperature range | -50.0 to 199.9 °C |
| Temperature accuracy | ±0.1 °C ±1 digit (meter only; total error includes the probe error: in 0 to 100 °C ±0.8 °C for a Pt-1000B probe, ±0.35 °C for a Pt-1000A probe) |
| Temperature probe (in set) | CT2B-121 with Pt-1000B resistor |
| Display | Large LCD, simultaneous readout of measured function and temperature |
| Memory | 50 results |
| Power supply | 12 V / 100 mA power adapter |
| Dimensions | L = 200 mm; W = 180 mm; H = 20/50 mm |
| Weight | 520 g |
| Warranty (meter) | 24 months |
Your email address will not published. Required fieled are marked
Other related products explored by visitors
The maximum purchase for this product is () pieces.