The Lutron MG-3002 is a wide-range, general-purpose AC/DC magnetic meter with a separate Hall probe and automatic temperature compensation. It reads in mT or G, DC up to 3,000.0 mT and AC up to 1,500.0 mT, shows N/S pole, and offers data hold, max/min recall and RS232/USB output.
The Lutron MG-3002 is a wide-range, general-purpose magnetic meter for DC and AC magnetic field measurement in industrial, mechanical, material, electrical and laboratory field use. Readings are shown in milli Tesla (mT) or Gauss (G) on a dual-function LCD.
The DC function covers 300.00 mT and 3,000.0 mT ranges (3,000.0 G and 30,000 G), and the AC function covers 150.00 mT and 1,500.0 mT ranges (1,500.0 G and 15,000 G) at 50 Hz / 60 Hz. Accuracy is ± (5 % rdg. + 10 digit) for DC and ± (5 % rdg. + 20 digit) for AC. The separate probe uses a uniaxial Hall sensor with a temperature sensor built into the probe head for automatic temperature compensation.
The meter runs on a 006P DC 9V battery or a DC 9V adapter input, and its RS232/USB PC serial interface allows connection to a computer through the optional cables.
| Manufacturer | Lutron Electronic |
|---|---|
| Model | MG-3002 |
| Circuit | Custom one-chip of microprocessor LSI circuit |
| Display | Dual function LCD, LCD size 52 mm x 38 mm |
| Measurement Units | mT (milli Tesla), G (Gauss) |
| Measurement Range, DC (mT) | Range 1: 300.00 mT x 0.01 mT; Range 2: 3,000.0 mT x 0.1 mT |
| Measurement Range, DC (G) | Range 1: 3,000.0 G x 0.1 G; Range 2: 30,000 G x 1 G |
| Measurement Range, AC (mT) | Range 1: 150.00 mT x 0.01 mT; Range 2: 1,500.0 mT x 0.1 mT |
| Measurement Range, AC (G) | Range 1: 1,500.0 G x 0.1 G; Range 2: 15,000 G x 1 G |
| Resolution | 0.01 / 0.1 mT, 0.1 / 1 G |
| Accuracy, DC | ± (5 % rdg. + 10 digit) |
| Accuracy, AC | ± (5 % rdg. + 20 digit) |
| Frequency Response (AC measurement) | 50 Hz / 60 Hz |
| Sensor | Hall sensor |
| Field Direction | Uniaxial |
| Temperature Compensation | Automatic (ATC); temperature sensor built into the probe head |
| Data Hold | Freezes the display reading |
| Memory Recall | Maximum and minimum value |
| Display Sampling Time | Approx. 1 second |
| Power Off | Manual off by push button or auto shut off |
| Data Output | RS232/USB PC serial interface (optional RS232 cable UPCB-02 or USB cable USB-01) |
| Operating Temperature | 0 to 50 °C |
| Operating Humidity | Less than 85 % R.H. |
| Power Supply | 006P DC 9V battery (Alkaline or Heavy duty type); DC 9V adapter input (AC/DC power adapter is optional) |
| Power Current | Approx. DC 20.5 mA |
| Standards | ISO-9001, CE, IEC1010 |
| Dimensions | Main instrument: 200 x 68 x 30 mm (7.9 x 2.7 x 1.2 inch); Probe: 198 x 25 x 19 mm (7.8 x 1.0 x 0.7 inch) |
| Weight | Approx. 270 g / 0.60 LB (including battery and probe) |