The Lutron AM-4210 is a portable vane anemometer with a separate probe that shows air velocity (0.8 to 30.0 m/s) and air temperature together on a dual display. It also works as a type K/J thermocouple thermometer and logs up to 1,600 readings, with RS232 output.
The Lutron AM-4210 is a portable anemometer with a separate vane probe. Its dual function LCD shows the air velocity and the air temperature at the same time, and the low-friction ball-bearing vane wheel is accurate at both high and low velocities. Air velocity is measured from 0.8 to 30.0 m/s and can be displayed in five units: m/s, km/h, ft/min, knots and mile/h.
The meter is a 2 in 1 instrument: besides the anemometer it is a thermocouple thermometer that accepts type K or type J temperature probes with 0.1 degree resolution, in °C or °F. The thermocouple probes are optional accessories.
A manual and automatic data logger with selectable sampling time stores up to 1,600 readings, and the RS232 PC serial interface sends the data to a computer. The meter runs on a 006P DC 9V battery or a DC 9V adapter.
| Manufacturer | Lutron Electronic |
|---|---|
| Model | AM-4210 |
| Functions | Anemometer; thermocouple thermometer (type K/J) |
| Air velocity (m/s) | Range 0.8 - 30.0 m/s; resolution 0.1 m/s; accuracy ± (2% + 0.2 m/s) |
| Air velocity (km/h) | Range 2.8 - 108 km/h; resolution 0.1 km/h; accuracy ± (2% + 0.8 km/h) |
| Air velocity (mile/h) | Range 1.8 - 67.0 mile/h; resolution 0.1 mile/h; accuracy ± (2% + 0.4 mile/h) |
| Air velocity (knots) | Range 1.6 - 58 knots; resolution 0.1 knots; accuracy ± (2% + 0.4 knots) |
| Air velocity (ft/min) | Range 160 - 5900 ft/min; resolution 1 ft/min; accuracy ± (2% + 40 ft/min) |
| Air temperature range | 0 °C to 50 °C, 32 °F to 122 °F |
| Air temperature resolution | 0.1 degree |
| Air temperature accuracy | °C: 0.8 °C; °F: 1.5 °F |
| Thermocouple thermometer resolution | 0.1 °C / 0.1 °F (type K and type J) |
| Type K range and accuracy (°C) | -50.0 to 1300.0 °C: ± (0.4 % + 0.8 °C); -50.1 to -199.9 °C: ± (0.4 % + 1 °C) |
| Type K range and accuracy (°F) | -58.0 to 2372.0 °F: ± (0.4 % + 1.5 °F); -58.1 to -327.8 °F: ± (0.4 % + 1.8 °F) |
| Type J range and accuracy (°F) | -58.0 to 2012.0 °F: ± (0.4 % + 1.5 °F); -58.1 to -327.8 °F: ± (0.4 % + 1.8 °F) |
| Type J range and accuracy below -50 °C | -50.1 to -199.9 °C: ± (0.4 % + 1 °C) |
| Thermocouple accuracy note | Accuracy value is specified for the meter only; type K probes are optional accessories |
| Thermocouple compensation | Automatic cold junction temperature compensation for type K/J; linear compensation for the full range |
| Sensors | Air velocity: conventional twisted vane arms and low-friction ball-bearing design; temperature: precision thermistor |
| Circuit | Custom one-chip microprocessor LSI circuit |
| Display | Dual function LCD, 44 mm x 29 mm |
| Data logger | Max. 1,600 points; manual (one record per button push) or auto sampling time 2, 5, 10, 30, 60, 600, 1800, 3600 seconds |
| Display sampling time | Approx. 1 second |
| Data hold / memory recall | Freezes the display reading; maximum and minimum value recall |
| Thermometer input socket | Standard two pins thermocouple socket |
| Data output | RS 232 PC serial interface |
| Power off | Auto shut off to save battery life, or manual off by push button |
| Operating temperature | 0 to 50 °C |
| Operating humidity | Less than 80% R.H. |
| Power supply | 006P DC 9V battery (alkaline or heavy duty type); DC 9V adapter input (AC/DC power adapter is optional) |
| Power current | Anemometer: approx. DC 10.5 mA; thermocouple thermometer: approx. DC 6.8 mA |
| Standards | ISO-9001, CE, IEC1010 |
| Dimensions | Meter: 135 x 60 x 33 mm (5.3 x 2.4 x 1.3 inch); probe sensor head: round, 72 mm dia. |
| Weight | 322 g / 0.71 LB (including probe and battery) |
Your email address will not published. Required fieled are marked
Other related products explored by visitors
The maximum purchase for this product is () pieces.